Hemodynamic and clinical impact of prosthesis–patient mismatch in the aortic valve position and its prevention
Philippe Pibarot and Jean G Dumesnil
Table 1
Table 1
Three Easy Steps to Avoid Prosthesis–Patient Mismatch
Step I: Calculate the patient’s body surface (BSA) area using the formula: |
BSA = ([Weightkg]0.425 × [heightcm]0.725) × 0.007184 |
Step II: Determine the minimal requirement for prosthetic valve effective orifice area (EOA) to avoid prosthesis–patient mismatch. |
Patient BSA (m2) | Minimal Valve EOA (cm2) for Indexed EOA >0.85 cm2/m2 (Ideal) | Minimal Valve EOA (cm2) for Indexed EOA >0.80 cm2/m2 | Minimal Valve EOA (cm2) for Indexed EOA >0.75 cm2/m2 |
---|---|---|---|
1.30 | 1.11 | 1.04 | 0.98 |
1.35 | 1.15 | 1.08 | 1.01 |
1.40 | 1.20 | 1.12 | 1.05 |
1.45 | 1.23 | 1.16 | 1.09 |
1.50 | 1.28 | 1.20 | 1.13 |
1.55 | 1.32 | 1.24 | 1.16 |
1.60 | 1.36 | 1.28 | 1.20 |
1.65 | 1.40 | 1.32 | 1.24 |
1.70 | 1.45 | 1.36 | 1.28 |
1.75 | 1.49 | 1.40 | 1.31 |
1.80 | 1.53 | 1.44 | 1.35 |
1.85 | 1.57 | 1.48 | 1.39 |
1.90 | 1.62 | 1.52 | 1.43 |
1.95 | 1.66 | 1.56 | 1.46 |
2.00 | 1.70 | 1.60 | 1.50 |
2.05 | 1.74 | 1.64 | 1.54 |
2.10 | 1.79 | 1.68 | 1.58 |
2.15 | 1.83 | 1.72 | 1.61 |
2.20 | 1.87 | 1.76 | 1.65 |
2.25 | 1.91 | 1.80 | 1.69 |
2.30 | 1.96 | 1.84 | 1.73 |
2.35 | 2.00 | 1.88 | 1.76 |
2.40 | 2.04 | 1.92 | 1.80 |
2.45 | 2.08 | 1.96 | 1.84 |
2.50 | 2.13 | 2.00 | 1.88 |
Step III: Choose a prosthesis using reference values for EOA of different types and sizes of prostheses (see Table 2). |